Declare the script language used and the source where the script functions are located. Type J Thermocouples -- coefficients of approximate inverse functions giving temperature, t90 , as a function of the thermoelectric voltage, E, in selected temperature and voltage ranges.  The functions are of the form:

t90 = d0 + d1E + d2E2 + ... dnEn

where E is in mV and t90 is in °C.



Temperature Range :    Coefficients :   

Download Tables of Thermoelectric Voltages and Coefficients

Temperature Range: -210 °C to 0 °C 0 °C to 760 °C 760 °C to 1200 °C Voltage Range: -8.095 mV to 0 mV 0 mV to 42.919 mV 42.919 mV to 69.553 mV d0 = 0.0000000 x100 0.000000 x100 -3.11358187 x103 d1 = 1.9528268 x101 1.978425 x101 3.00543684 x102 d2 = -1.2286185 x100 -2.001204 x10-1 -9.94773230 x100 d3 = -1.0752178 x100 1.036969 x10-2 1.70276630 x10-1 d4 = -5.9086933 x10-1 -2.549687 x10-4 -1.43033468 x10-3 d5 = -1.7256713 x10-1 3.585153 x10-6 4.73886084 x10-6 d6 = -2.8131513 x10-2 -5.344285 x10-8   d7 = -2.3963370 x10-3 5.099890 x10-10   d8 = -8.3823321 x10-5   Error Range: -0.05 °C to 0.03 °C -0.04 °C to 0.04 °C -0.04 °C to 0.03 °C


Temperature Range :    Coefficients :   

Download Tables of Thermoelectric Voltages and Coefficients