Declare the script language used and the source where the script functions are located. Type K Thermocouples -- coefficients of approximate inverse functions giving temperature, t90 , as a function of the thermoelectric voltage, E, in selected temperature and voltage ranges.  The functions are of the form:

t90 = d0 + d1E + d2E2 + ... dnEn

where E is in mV and t90 is in °C.



Temperature Range :    Coefficients :   

Download Tables of Thermoelectric Voltages and Coefficients

Temperature Range: -200 °C to 0 °C 0 °C to 500 °C 500 °C to 1372 °C Voltage Range: -5.891 mV to 0 mV 0 mV to 20.644 mV 20.644 mV to 54.886 mV d0 = 0.000000 x100 0.000000 x100 -1.318058 x102 d1 = 2.5173462 x101 2.508355 x101 4.830222 x101 d2 = -1.1662878 x100 7.860106 x10-2 -1.646031 x100 d3 = -1.0833638 x100 -2.503131 x10-1 5.464731 x10-2 d4 = -8.9773540 x10-1 8.315270 x10-2 -9.650715 x10-4 d5 = -3.7342377 x10-1 -1.228034 x10-2 8.802193 x10-6 d6 = -8.6632643 x10-2 9.804036 x10-4 -3.110810 x10-8 d7 = -1.0450598 x10-2 -4.413030 x10-5   d8 = -5.1920577 x10-4 1.057734 x10-6   d9 =   -1.052755 x10-8   Error Range: -0.02 °C to 0.04 °C -0.05 °C to 0.04 °C -0.05 °C to 0.06 °C


Temperature Range :    Coefficients :   

Download Tables of Thermoelectric Voltages and Coefficients