Declare the script language used and the source where the script functions are located. Type E Thermocouples -- coefficients of approximate inverse functions giving temperature, t90 , as a function of the thermoelectric voltage, E, in selected temperature and voltage ranges.  The functions are of the form:

t90 = d0 + d1E + d2E2 + ... dnEn

where E is in mV and t90 is in °C.



Temperature Range :    Coefficients :   

Download Tables of Thermoelectric Voltages and Coefficients

  -200 °C to 0 °C 0 °C to 1000 °C Voltage Range: -8.825 mV to 0 mV 0 mV to 76.373 mV d0 = 0.0000000 x100 0.0000000 x100 d1 = 1.6977288 x101 1.7057035 x101 d2 = -4.3514970 x10-1 -2.3301759 x10-1 d3 = -1.5859697 x10-1 6.5435585 x10-3 d4 = -9.2502871 x10-2 -7.3562749 x10-5 d5 = -2.6084314 x10-2 -1.7896001 x10-6 d6 = -4.1360199 x10-3 8.4036165 x10-8 d7 = -3.4034030 x10-4 -1.3735879 x10-9 d8 = -1.1564890 x10-5 1.0629823 x10-11 d9 =   -3.2447087 x10-14 Error Range: -0.01 °C to 0.03 °C -0.02 °C to 0.02 °C


Temperature Range :    Coefficients :   

Download Tables of Thermoelectric Voltages and Coefficients