Overall Energy Resolution (eV):
                            Calibration:
                            Au,Cu = 84.00,932.67
Charge Reference:
                            Adventitious carbon
Energy Scale Evaluation:
                            Reliable, with one-point correction of energy scale
Comment:
                            The composition determined by XPS was K0.467Br0.496. Peak locations: Voigt function. The intensity ratio of the Gaussian/Lorentzian components was 80/20. An electron flood gun (Ep = ~3 eV, Ip = 1 mA) and a Ni mesh were used for charge compensation. FAT mode. The spectra were recorded at normal emission.
 
                            cleaved crystal, crystal, solid on double-sided tape
 
                        Method of Determining Specimen Composition:
                            X-ray Photoelectron Spectroscopy
Method of Determining Specimen Crystallinity:
                            Specimen Temperature (K):
                            320