Overall Energy Resolution (eV):
Calibration:
Au,Cu = 84.00,932.67
Charge Reference:
Adventitious carbon
Energy Scale Evaluation:
Reliable, with one-point correction of energy scale
Comment:
The composition determined by XPS was K0.467Br0.496. Peak locations: Voigt function. The intensity ratio of the Gaussian/Lorentzian components was 80/20. An electron flood gun (Ep = ~3 eV, Ip = 1 mA) and a Ni mesh were used for charge compensation. FAT mode. The spectra were recorded at normal emission.
cleaved crystal, crystal, solid on double-sided tape
Method of Determining Specimen Composition:
X-ray Photoelectron Spectroscopy
Method of Determining Specimen Crystallinity:
Specimen Temperature (K):
320